When checking a humidity reading, also look up the range, accuracy conditions, response time and operating environment. We use the Tanita TT-559/572 household meter and the HygroVue 5 observation sensor to illustrate how to read specifications. For mountain use, check the operating conditions of the model you actually have.

Display range and rated accuracy cover different things

The Tanita TT-559/572 digital thermometer and hygrometer shows temperatures from −5.0 to 50.0 °C and humidity from 20% to 95%. Outside the temperature range it shows codes such as LL.L or HH.H. The manual says humidity may not be measured accurately when the temperature shows either code.

Temperature and humidity have separate readouts, and their operating conditions can be linked. On these models, out-of-range temperatures may prevent accurate humidity readings and trigger a different display. Check temperature and error codes as well as humidity and leave missing values unestimated.

The Campbell Scientific HygroVue 5 manual (revision 07/2026) gives an operating range of −40 to 70 °C for temperature and 0% to 100% relative humidity (RH). Range and display behavior can vary widely among devices that measure temperature and humidity.

The range an instrument can display is different from the range where it meets a particular accuracy rating. The TT-559/572 can display 20% to 95% humidity, and its tightest accuracy band covers only part of that range. A number on the display does not guarantee high accuracy. The two instruments have different purposes and designs, so compare each with its own intended use.

Accuracy comes in bands with conditions attached

The TT-559/572 is rated ±1 °C from 0 to 40 °C and ±2 °C outside that band. Humidity accuracy is ±5 percentage points from 35% to 75% RH and ±10 points elsewhere in the range. Those tolerances cannot guarantee the error if the device is used outside its specifications, has aged or is wet. The display alone cannot always establish the true humidity.

For the HygroVue 5 at 25 °C, humidity accuracy is ±1.8 percentage points from 0% to 80% RH and ±3 points from 80% to 100% RH. The table also gives temperature dependence below ±1 point RH from −40 to 60 °C. Read the accuracy at the reference temperature, temperature dependence and long-term stability as separate specifications. A single figure does not establish low-temperature accuracy.

The manual lists short-term hysteresis of less than ±1 percentage point RH. It also notes that temperature and relative-humidity accuracy figures are relative to factory standards. Check the footnotes for the conditions and reference behind the numbers.

Two different readings do not automatically indicate a fault. If both devices actually meet ±5-point accuracy in the same environment, they can differ by up to 10 percentage points. In practice, also check placement, response lag and wear. That calculation alone cannot diagnose a device.

Update intervals and response times affect how current the number is

The TT-559/572 measures about once every 20 seconds. Its manual says it takes about 30 minutes after setup to display a correct value. Refresh interval and settling time are different. An immediate reading need not be ready to use, and 30 minutes does not guarantee accuracy under every condition. Follow placement instructions, including keeping it away from direct sunlight and air blowing from appliances.

The HygroVue 5’s 63% response time in airflow of 1 m/s is 130 seconds for temperature and 8 seconds for humidity. The humidity value is for a fitted filter and 25 °C. At 63%, a little over half of the change has registered. The reading has not fully settled. These specifications illustrate how response differs between temperature and humidity on this sensor.

To watch the display at home, use the device as directed and note placement time and reading time. A steady number may reflect rounding or the refresh interval, and positioning can still bias it. It does not prove accuracy. On the mountain, stop somewhere safe and avoid waiting in cold or bad weather just to measure.

Read separately about moisture, condensation and storage

The TT-559/572 is not waterproof. The instructions say to avoid storing or using it in damp or splash-prone places, or using it with wet hands. Keep it out of direct sun and airflow from air conditioners or humidifiers. Avoid storage by heaters or in dusty places or with large swings in temperature or humidity. This is an indoor instrument with specific limits, even though it measures humidity.

The HygroVue 5 installation guide says to protect the tip filter from liquid water and use a radiation shield outdoors. Measuring humidity does not make a sensor fit for direct contact with water droplets. Check the guidance for each part and how to install it, as the housing rating need not apply to the sensing element.

The Japan Meteorological Agency also uses a ventilated shield and a fan to protect its instruments from precipitation and sun. The Okinawa Meteorological Observatory explains that its hygrometer measures changes in capacitance in a moisture-absorbing polymer film. This describes an official station. Check the actual design and specifications of your own instrument.

Sensirion’s SHTxx document mainly covers handling sensor components during production and storage. It describes storage conditions and warns that chemical exposure can cause drift, including during field use. A specification based on clean air leaves room for chemical exposure to affect the sensor outdoors. Follow the finished device’s manual instead of applying the component’s storage limits or reconditioning procedures.

In short: copy six details from your manual to put readings in context

Before leaving, copy six details from your own manual to put each humidity reading in context. First, note the display range and how the device shows out-of-range readings. Second, record accuracy bands and conditions such as temperature and reference. Third, note the refresh interval and settling time after setup. Fourth, note the response time and test conditions such as airflow and filter. Fifth, list unsuitable places to install or use it. Sixth, list unsuitable storage conditions. Leave information absent from the manual blank and mark it unconfirmed.

Use the sheet to check whether the reading was in range, whether the device had settled and whether wetness or condensation was possible. Treat readings as approximate when many details are missing. Factor in the sky, wind and how you feel as well as humidity, leaving room to stop or turn back while it is still safe.